Methodology edit

Mar 23, 2020

Leave a message

Methodology edit 
Speaking of the measurement of the permeability μ, it seems very simple. Just wind a few turns of the coil on the material sample ring, measure its inductance, find a formula and finish it. In fact, it is not true that for the same sample ring, different instruments, different numbers of turns, different voltages, or different frequencies may be used to measure far-different permeability. The reason for the great difference in test results is that not every tester has the energy to figure it out. This article mainly discusses the influence of different test turns and calculation formulas on permeability measurement. 
Impact of calculation formula 
As you know, the method of measuring the permeability μ is generally to measure the inductance L of N turns of the coil on the sample ring. For a ring core with a small inner diameter, the inner diameter is not as easy to measure as the wall thickness. The path length is taken as the magnetic path length of the magnetic core. The magnetic permeability calculated using them is called the ring permeability of the material. Some people say that the permeability measured with a ring sample is called the ring permeability, and this statement is incorrect. In fact, the magnetic permeability of the ring is higher than the true permeability of the material, and the thicker the wall of the sample ring, the greater the error. 
For the sample ring, the magnetization field is not uniform in the radial direction when the magnetomotive force is excited with the same number of turns. The closer to the outer side of the ring wall, the weaker the magnetic field. Under the condition that the magnetic permeability μ is constant throughout the sample ring, the closer to the outside of the ring wall, the lower the magnetic flux density B of the ring. In order to eliminate the influence of this uneven magnetization on the measurement, we consider the sample ring to consist of an infinite number of thin-walled rings with radius r and infinitely thin wall thickness dr. 
If the sample ring is composed of the same material, the calculated permeability is the true permeability μ of the material. It is slightly lower than its ring permeability. 
Effect of test coil turns N 
Since the inductance L is directly proportional to the number of turns N2, the calculated permeability μ should no longer be related to the number of turns N, but in fact it is often related. 
Regarding the measurement of material permeability, the test frequency used is generally not high, and it is often tested at a frequency of 1kHz or 10kHz. The test signal is generally a sinusoidal signal, because the frequency is not high, the resistance part of the coil resistance of the sample surround group can be ignored, and the winding coil is regarded as a pure inductance L and connected to the measuring instrument. The test equivalent circuit is shown in the figure. The effective value of the voltage generated by the signal source of the instrument is U, and Ri is the output impedance of the signal source. 
When measuring the permeability, the magnetization field strength in the sample ring is related to the number of turns of the test coil. When the number of turns is a certain value, the magnetic field strength will reach the strongest value. The magnetic permeability of the material is closely related to the magnetizing field strength, so the measurement of the magnetic permeability is related to the number of test coil turns. The effect of the number of turns on the permeability test will be discussed in detail with reference to the figure.

For more information,please contact us at sales@xfullstar.com